PRESSURE TRANSMITTER

BV9000 BAROMETRIC/VACUUM PRESSURE TRANSMITTER

•  Dual Range— barometric or vacuum

•  Compatible with Aquistar® Dataloggers

•  Quick-connect vacuum port

•  Thermally stable

DESCRIPTION

The BV9000 is a barometric/vacuum pressure transmitter designed to provide accurate barometric and vacuum pressure measurements. A three-wire, 4-20 mA output device, the sensor features vacuum and barometric ranges of 850-1100 mbar (12.3 to 16.0 PSIA) and 600-1100 mbar (8.7 to 16.0 PSIA).

OPERATION

The BV9000 requires a datalogger or control system and 10.5-20 VDC excitation for operation; it achieves stability after a one-second warm-up time. The sensor comes equipped with an inlet quick-connect fitting that allows it to be easily attached to vacuum chambers, well seals or pipe systems.

PRIMARY APPLICATIONS

INW’s BV-9000 barometric/vacuum sensor may be used to:

•   Supplement aquifer test data in leaky or confined conditions.

•   Measure vacuum pressure during vapor extraction pilot testing

•   Be used to barometrically compensate absolute pressure sensors for level measurement.

BV9000 barometric/vacuum pressure transmitter’s internal components.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

BV9000 BAROMETRIC/VACUUM

PRESSURE TRANSMITTER

Part #

DESCRIPTION

5773C800

BV9000 Barometric / Vacuum Pressure Transmitter

 

BV 9000 CABLE OPTIONS

Part #

Description

5776E508

Non-Vented PU Cable (Sold per Foot.  Please specify length)

5776E400

M6 Connector

 

MISCELLANEOUS

Part #

DESCRIPTION

5776D250

Aquastar® DL-2 Data Collection System w/ Aqua4 Software

DIMENSIONS AND SPECIFICATIONS

HOW TO ORDER

• Specify cable length

• Choose appropriate options

 

MECHANICAL

Transmitter

Body Material

ABS

Wire Seal Materials

Nylon & Neoprene

Inlet QC Materials

Delrin® & Buna-N

Weight

.75 lbs

Cable

Conductor Type

6-Conductor

OD

0.28” max.

Break Strength

138 lbs.

Maximum Length

2000 feet

Weight

4 lbs. per 100 feet

ELECTRICAL1

Non-linearity2

±1.2%  (max.)

±0.45%  (typical)

Thermal Error3

(0-50 C Reference 25° C)

±3.0% FSO (max.)

±1.8% FSO (typical)

Repeatability

±3.0% FSO (max.)

±1.8% FSO (typical)

Hysteresis

±0.03% FSO (max.)

±0.009% FSO (typical)

Long-Term Stability4

±0.75% FSO (max.)

±0.25% FSO (typical)

Transmitter Operating Voltage Requirement

10.5 - 20 VDC)

Operating Temperature

-30 - 100° C

Range Output Current 4 - 20 mA

1 These applications are based on INW’s standard range

of  600 - 1100 mbar.  Other ranges will vary proportionate

to the range of calibration.

2 Best fit straight line method.

3 0-50°C range.

4 Based on 1000 pressure cycles and 30 temperature

cycles over a one-week period normalized to 25°C.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

DRILLERS SERVICE, INC.                      HICKORY, NC USA                    800-334-2308,  EXT. 230